Diffusion Furnace Instructions

Things to have ready:
Quartz boat, with wafers loaded evenly spaced
Quartz boat holder
High Temperature gloves

Procedure:
Things to turn on:
The power light should be on
and the power output gauges should be at100% (5)output

Set three Temp dials to 800
Wait for furnace to reach desired temperature
While waiting check water level in bubbler at back of furnace, fill as necessary
Turn power gauge on bubbler heater past 20 then down to 15
Once temperature reached set power dials so that all outputs are 20-50%
Turn on Oxygen at turret
Switch Furnace valve from nitrogen to oxygen
Turn off
Nitrogen at wall turret



Take cap off end of quartz tube
Use quartz boat holder to place quartz boat in opening of Quartz tube
Use glass rod to push quartz boat into tube opening (ONLY PUSH BOAT IN ENOUGH TO BE ABLE TO REPLACE THE END CAP, pushing boat in to far to fast will break all the wafers)
Replace end cap
Remove small plug held onto end cap with metal clamp
Use glass rod to push quartz wafer boat 5 inches into the furnace every 1 minute
Once boat at center of furnace (When pushing rod reaches tape mark nearest furnace) remove rod and replace small plug
Wait for necessary time
Remove plug on end cap
Use glass rod to remove wafer boat 5 inches out of the furnace every 1 minute
Once boat at end of tube remove glass rod, replace small plug into endcap
Remove endcap and place down on table
Place wafer boat holder in front of tube opening
Use glass rod to pull boat into boat holder
Replace endcap and allow wafers to cool under hood for 10 minutes
Turn off Control panel power and heater element power
Turn off Oxygen supply and bubbler heater
Turn On Nitrogen supply
Allow nitrogen to flow through tube until the tube cools to 200 C
Once tube cooled turn off Nitrogen and cooling water
procedure prepared by Francisco Tejada, Spring
2003; modified by La Vida Cooper, Spring 2003.